|
|
1. To monitor and validate the cooling point of
sterilization unit
|
[mode 1]
apply the thermocouple, chemical indicator, EO-gas test tube to check,
monitor
and validate
the cooling point of sterilized water, liquid, and products at the
steam,
autoclave, dry heat, EO-gas and radiation sterilization unit
[mode 2]
apply the self contained biological indicator, spores strip to check,
monitor and
validate the
cooling point of sterilized water, liquid, and products at the steam,
autoclave,
dry heat, EO-gas and radiation sterilization unit
[mode 3]
apply the process chemical indicator to check, monitor and validate the
cooling
point of
sterilized water, liquid, and products at the autoclave, steam sterilizer
|
2. To monitor and validate sterilization effect of
sterilized water, liquid
and products at the cooling
point sterilization unit
|
1. to check, monitor and validate
the sterilization temperature, concentration
and dosage of cooling point up to
expected sterilization temperature, dosage
and concentration or not
apply the thermocouple, chemical indicator,
EO-gas test tube to check, monitor and
validate the cooling point of sterilization
temperature, dosage and concentration up to
expected sterilization temperature, dosage and concentration or not
2. to check, monitor and validate
the sterilization rate of sterilization effect
being acted on the cooling point of
sterilized water, liquid and products
a. can be applied the self contained
biological indicator, or spores strip biological
indicator to monitor and validate the sterilization rate of cooling point
b. can also be applied the sterilization
temperature, dosage, concentration and times
being acted on the cooling point of sterilized water, liquid and products
3. to check, monitor and validate
the sterilization times after the cooling point
of
sterilized water, liquid and products being functioned up to the expected
sterilization
temperature, dosage and concentration
a. can be applied process chemical indicator
to
monitor and validate the sterilization
times after the cooling point of autoclave being functioned up to
expected
sterilization
temperature, dosage, concentration
b. can also be applied the thermocouple,
chemical indocator, EO-gas test tube to check
when the cooling
point of sterilization temperature, dosage, concentration are
functioned up to expected
sterilization temperature, dosage, concentration to monitor
and validate the correct sterilization times
|
三. After sterilization, the correct sterilization unit
operation of aseptic
treatment, monitoring and validation to monitor, validate and
confirm
the sterilization effect of
the sterilization validation indicator
|
1. After sterilization, the
sterilization validation indicator are shown that the
following
different sterilization effect
a. The
monitoring and validation of thermocouple, temperature chemical indicator, EO-
gas test tube, radiation dosage
chemical indicator
only to monitor and validate the
sterilization temperature, dosage, concentration up to
expected
sterilization temperature, dosage, concentration or not, can not monitor and
validate the sterilization rate,
and the bacteria residues up to aseptic free or not
b. The
monitoring and validation of spores strip and self contained biological
indicator
only to monitor and validate the
lethal rate of sterilization effect after sterilization, but
can not monitor and validate the
clear bacterial residues of sterilized water, liquid and
products up to aseptic free or not
1) 105
CFU/ml biological indicator to monitor and validate 5D(99.999%) lethal rate
2) 106
CFU/ml biological indicator to monitor
and validate 6D(99.9999%) lethal rate
c. The
monitoring and validation of process chemical indicator
only
applied to autoclave (steam) sterilization application which can be monitored
and
validated
the sterilization effect of sterilization temperature and times up to 121 o
C, 30 minutes, or 131 o C,
30 minutes, to do 8D (99.999999%) sterilization rate, but can not
monitor
and validate the clear bacterial residues of sterilized water, liquid and
products
up to aseptic free or not
d. The
monitoring and validation of radiation dosage chemical indicator:
can be
directly monitored and validated the sterilization dosage and times being acted
on the
cooling point of sterilized water, liquid and products for expected
sterilization
rate of
sterilization effect application |
2. After sterilization, the
aseptic monitoring and validation of the unit operation
of
biological indicator
| self contained biological indicator
|
spores strip of biological indicator |
|
|
四. After sterilization, the correct sterilization unit
operation of aseptic
treatment, monitoring and validation to monitor, validate and
confirm
the bacterial residues of
sterilized water, liquid and products low than
≦10-6
CFU/ml, up to aseptic free or not
|
After sterilization, The
aseptic monitoring and validation of bacterial residues
of sterilized water, liquid
and products low than≦10-6
CFU/ml, up to aseptic
free are not determined by
the single aseptic monitoring and validation unit
operation of biological
indicator, which are determined by the pre-sterilization
factor of the bacterial
contaminations of pre-sterilization water, liquid, products
to be purified up to low
than≦100 CFU/ml, and the process sterilization factor
to be proceeded the 8 D
(99.999999%) sterilization rate or not
the conventional aseptic
monitoring and validation of biological indicator are
only applied to monitor, validate
and confirm the process sterilization factor up
to 5 D (99.999%) , and hard to
confirm the bacterial residues of post- sterilized
water, liquid and products low
than≦10-6 CFU/ml, up to aseptic
free
The aseptic monitoring and
validation of bacterial residues of post-sterilized
water, liquid and products
are low than≦10-6
CFU/ml up to aseptic free or not
which can be performed the
following way
[mode 1] to establish aseptic monitoring and
validation unit operation to monitor and
validate the lethal
rate of cooling point biological indicator higher than the lethal
rate from the
pre-sterilization bacterial contaminations down to the post-
sterilization
bacterial residues low than≦10-6
CFU/ml
[mode 2] to establish aseptic monitoring and
validation unit operation to monitor and
validate the
cooling point process sterilization rate higher than the lethal rate
from the
pre-sterilization bacterial contaminations down to the post-sterilization
bacterial residues
low than≦10-6 CFU/ml
[mode 3] can be monitored and validated by
the LAL endotoxin test and IPT pyrogen test
a) LAL endotoxin
test: to monitor and validate the endotoxin content low than
≦0.2 EU/ml or not, if low than≦0.2
EU/ml, the residues
of gram-negative bacteria will be low than≦10-6
CFU/ml
b) IPT pyrogen
test: to monitor and validate the pyrogen content low than≦0.2
EU/ml or not, if low than≦0.2 EU/ml, the
residues of gram-
positive bacteria,
gram-negative bacteria, virus, fungi will be
low than≦10-6
CFU/ml, up to aseptic
free |
[The analysis of biological indicator are applied to
monitor and validate
the bacterial
residues low than≦10-6
CFU/ml, up to aseptic free]
|
1. The standard sterilization application of US FDA.
AAMI, Europe ESH,
EP, to
sterilize the bacterial residues of
post-sterilization water, liquid and products
up to low than≦10-6
CFU/ml, up to aseptic free
[point] The bacterial contaminations of
pre-sterilization water, liquid and products are easy
to
be purified up to bacteria free, the bacteria count are reduced down to low than
≦1 CFU/ml (≦10
0 CFU/ml), why are only setted up low than≦100 CFU/ml, 2. The conventional sterilization process operation of aseptic
treatment, monitoring
and validation of autoclave, steam, dry heat,
EO-gas and radiation sterilization
are applied at hospitals, medical device,
pharmaceutical, biological products
3. The correct sterilization process operation of aseptic treatment,
monitoring and
validation of autoclave, steam, dry heat,
EO-gas
and radiation sterilization are
applied at hospitals, medical device,
pharmaceutical, biological products
4.
The
reason why the autoclave, steam, dry heat, EO-gas, radiation sterilization
application
of hospitals, pharmaceutical, medical device and biological products
are
needed to apply biological indicator as the sterilization effect of
monitoring and validation ?
why
the sterilization effect of steam, dry heat, EO-gas, radiation
sterilization
effect
are needed to apply the 10 6 CFU/ml biological indicator to
substitute the 10 5 CFU/ml biological indicator as sterilization monitoring and validation ?
1)
The
comparison of 10 6 CFU/ml biological indicator and 10 5
CFU/ml biological indicator are applied at the sterilization monitoring and validation
2)
The sterilization application mode of 10 6 CFU/ml biological
indicator and 10 5 CFU/ml
biological
indicator are compared to the standard sterilization application mode
5.
why
the LAL endotoxin test, pyrogen test, sterilization process chemical
indicator can be substituted the biological indicator to be acted
as the
test indicator of sterilization monitoring and validation ?
1)
the description of LAL endotoxin test are acted as the rapid sterilization
monitoring,
validation
and aseptic free indicator
2)
the description of pyrogen test are acted as the rapid sterilization
monitoring, validation
and
aseptic free indicator
3)
sterilization process chemical indicator can be substituted the biological
indicator to be
acted
as rapid sterilization monitoring and validation indicator
4)
the sterilization application mode of LAL endotoxin test, pyrogen
test and sterilization
process
chemical indicator are applied to acted as the rapid sterilization
monitoring,
validation
and aseptic free indicator
|
Bright Glory enterprise,Co.Ltd/
SanDa membrane &
Nano,Co.Ltd
5F-3,
No.162, Sec. 4, Chunghsiao E. Road, Taipei, Taiwan
Tel:
886-02-2731-8306
Fax: 886-02-2731-8221
web site: www.bge.com.tw web site: http://3w.bge.com.tw Email: hong@bge.com.tw